美国BC SA-2010S医用安规测试仪基本特征:
• LED状态指示灯
• 音频反馈
• 触摸控制键 - 无旋钮
• 10个通用患者导线输入
• 高冲击塑料外壳
SA-2010S安规测试仪安全分析仪功能:
• 线路电压测量
• 待测电流测量设备
• 接地/接地引线电阻
• 接地/接地漏电流
• 外壳/机箱漏电流
• 外部电阻
• 外部漏电流
• 源插座接线完整性监视器 •
真有效值测量
• AAMI ES1-1993或IEC 601可选测试负载
• 90至264 VAC运行
• 全20安培额定值
• 自检点
• 外部可更换接地保险丝
• 自动负载反转延迟
• 患者导致引线漏电流
• 患者导通接地/漏电流
• 患者隔离漏电流
• 外部隔离漏电流
患者模拟器功能: • 正弦:10,60和100HZ • 心电图:30,60,120和240 BPM • 方形:0.125和2.0 HZ • 三角形:2 HZ
• 高电平输出(1 V pp)
• 幅度精度:+/- 2%
• 频率精度:+/- 0.5%
美国BC SA-2010S-R安规测试仪:
• RS-232串行通信
The BC Biomedical Model SA-2010S Series is a microprocessor-based electrical safety analyzer with a built-in patient simulator. It allows for a multitude of tests to be performed on a device using the same unit and lead connections. It provides a full electrical safety analyzer, as well as ECG Simulation with four waveforms with constant QRS duration and six machine performance testing waveforms. There is Patient Lead testing for 10 inputs.
SA-2010S Series Basic Features:
• Led status indicators
• Audio feedback
• Touch control keys – no knobs
• 10 universal patient lead inputs
• High impact plastic case
Safety Analyzer Capabilities:
• Line voltage measurement
• Device under test current measurement
• Earth / ground lead resistance
• Earth / ground leakage current
• Enclosure / chassis leakage current
• External resistance
• External leakage current
• Source receptacle wiring integrity monitor
• True RMS measurements
• AAMI ES1-1993 or IEC 601 selectable test loads
• 90 to 264 VAC operation
• Full 20 amp rating
• Self test points
• Externally replaceable ground fuse
• Automatic load reversal delay
• Patient lead to lead leakage current
• Patient lead to earth / ground leakage current
• Patient isolation leakage current
• External isolation leakage current
Patient Simulator Capabilities:
• Sine: 10, 60 AND 100 HZ
• ECG: 30, 60, 120 AND 240 BPM
• Square: 0.125 AND 2.0 HZ
• Triangle: 2 HZ
• High level output (1 V p-p)
• Amplitude accuracy: +/- 2%
• Frequency accuracy: +/- 0.5%
SA-2010S-R:
• RS-232 serial communications